Henan Xinke Analytical Instruments Co.,Ltd
|
Price: | 1000.0 USD |
Payment Terms: | T/T,L/C |
Place of Origin: | Henan, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Industrial Analyzer
Product Name: Automatic industrial analyzer
Product Type: XKGF-3000
Scope:
Coal. electric power. chemical. geological survey. environmental protection. quarantine inspection. scientific research and other related industries and departments and colleges test materials such as coal. coke. moisture. ash content. volatile. and calculate the fixed carbon. hydrogen content and calorific value; To analyze the carbon content of fly ash. ash; To analyze the ignition loss of cement.
Features:
Automation and control
1. The furnace insulation materials: using a circular furnaces. furnace temperature distribution is more uniform. stable way to solve the assembly line frequently open oven door cause uneven and unstable technical problems.
2. Interface: CAN bus technology. the use of shielding and lightning protection circuit devices. anti-magnetic interference. anti-stimulus interference.
3. Temperature control mode: PID algorithm. anti-overshoot heating. safe. precise temperature control. heating speed.
4. Gas components: the use of low noise and piston pump. (Choose from nitrogen and oxygen)
5. The control mode: position control. More effective control of the host computer operation is not subject to unilateral causes uncontrollable. Fault alarm to protect themselves. And with the shift of the feature to make precise positioning technology with a unique micro-between.
6. Analysis and testing system: the vast majority of the company may be connected coal analysis equipment. You can automatically upload data. generate reports. (Users can design reports according to their needs).
7. Using pneumatic door slowly and automatically switch to avoid the cold and heat damage to the furnace.
Accurate and reliable data
1. A variety of testing methods for the user to arbitrarily set. there is the national standard arbitration. rapid method. it can also be set by the user definable.
2. Each can use the classic test ash gray slowly. in full compliance with the national standard in not less than 500 ℃ 30Min furnace temperature was raised requirements to ensure the accuracy of the ash content of the test.
3. Can be fused to the Quality Management System data and generate reports.
Technical parameters:
Sample weight: 0.8-1.2g. size: ≤0.2mm
Furnace temperature range: room temperature ~1000 ℃; Temperature accuracy: ± 1 ℃. Temperature control point: Water 105~110 ℃. Ash 815 ± 10 ℃. Volatile: 900 ± 10 ℃
Test time:<150Min / 20-like (moisture. ash. volatile matter and fixed carbon calculation)
Voltage: 220 ± 22VAC 50 ± 1HZ. maximum power:<5Kw
Accuracy: meet and better than the national standard
Dimensions (mm): 555 × 580 × 525